Selectra E
₨1,200,000.00
- Kinetic measurement with linearity check
- Bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
- Two point measurement
- Graphic plot of all measuring points
- Automatic rerun with sample reduction
- Non-linear calibration curves
- Description
- Reviews (0)
Description
Throughout
- Up to 180 tests per hour
- Up to 300 tests per hour with ISE unit
Reagent System
- Rotor with 24 positions for 25 ml bottles and 8 positions for 5 ml bottles. All positions can be assigned as R1, R2, R3. Adapters for 5 ml bottles in 25 ml positions
- 5 pairs of 25 ml positions can be used for 50 ml bottles
- Reagent 1 volume 110-400 µl
- Reagent 2 volume 0-180 µl
- Reagent 3 volume 0-180 µl
- Reagent disk compartment is cooled to approx. 12° C below ambient temperature
- Preheated reagent needle with level detection and integrated mixer
Sample System
- Sample rotor containing
- Outer segment with 51 positions for samples and/or controls
- Inner segment for calibrators, controls, stats, blank and wash solution
- Stat and pediatric functionality
- Continuous loading
- All positions can contain 5 ml primary tubes or sample cups
- Sample volume 1-30 µl per test, programmable in steps of 0.1µl
- Sample probe with level detection and intergrated mixer
Cuvette-Rotor
- Semi-Disposable rotor with 48 cuvettes. Path length 7mm
- > 10.000 tests per rotor
- Min. measuring volume 220 µl
- Measuring temperature 37° C, controlled by Peltier elements
Washing Unit
- Cuvette-washing with 4 X 500 µl of water. The unit is equipped with liquid sensors. Waste is seperated into diluted and concentrated (sample/reagent mixture and optional ISE unit) waste. Cuvettes are dried before use
Light Source
- Quarts-iodine lamp 12V-20W
Optics & Wavelength Range
- Optical unit with an 8 position filter wheel
- Automatic wavelength selection by 8-position filter wheel (340, 376, 405, 436, 505, 546, 578 and 620 nm)
- Other wavelengths avaiable on request
- Half bandwidth 8 to 12 nm
Photometric Range
- -0.1 to 3.0 Absorbance
Analytical Modes
- Kinetic measurement with linearity check
- Bichromatic end point measurement with or without bichromatic reagent blank and/or sample blank correction
- Two point measurement
- Graphic plot of all measuring points
- Automatic rerun with sample reduction
- Non-linear calibration curves